Case Report

Malposition of Central Venous Pressure Catheter as a Rare Cause of Tricuspid Valve Prosthesis Malfunction: A Case Report malfunction; a case report

Abstract

Prosthetic valve malfunction is a hazardous clinical entity that needs early diagnosis and management. Hereby, we present a case of tricuspid valve (TV) malfunction due to interaction of central venous pressure (CVP) catheter with TV leaflets. Our case was a 38-year-old male who referred to our center with the diagnosis of TV malfunction. Confirming the diagnosis of TV malfunction, intravenous Heparin was initiated. Anticoagulation therapy was discontinued early due to the development of significant headache in the setting of subarachnoid hemorrhage. Due to contraindication of anticoagulation therapy, he underwent redo TV. On the 4th post-operative day, echocardiography examination illustrated significant increase in gradients of TV prosthesis. At this time, fluoroscopy depicted interaction of CVP catheter with TV leaflets due to its cardiac advancement resulted in fixed TV leaflets in semi-open position. By extraction of CVP catheter the TV gradients improved immediately. Malposition of CVP catheter or large CVP catheter could be considered as one of the rare causes of TV malfunction. Repositioning of CVP catheter brings dramatic improvement in this case.

1. Cardoso G, Trabulo M, Andrade MJ, Ribeiras R, Rodrigues R, Neves JP, et al. A rare case of two mechanisms of prosthetic valve dysfunction in the same patient. Rev Port Cardiol 2013;32:1037-41.
2. Ma WG, Hou B, Abdurusul A, Gong DX, Tang Y, Chang Q, et al. Dysfunction of mechanical heart valve prosthesis: experience with surgical management in 48 patients. J Thorac Dis 2015;7:2321-9.
3. Shapira Y, Sagie A, Jortner R, Adler Y, Hirsch R. Thrombosis of bileaflet tricuspid valve prosthesis: clinical spectrum and the role of nonsurgical treatment. Am Heart J 1999;137:721-5.
4. Apfelbaum, Jeffrey L. Practice guidelines for central venous access 2020 an updated report by the American Society of Anesthesiologists Task Force on central venous access. Anesthesiology 2020;132.1:8-43.
5. McGee DC, Gould MK. Preventing complications of central venous catheterization. N Engl J Med 2003;348:1123-33.
6. Ruesch S, Walder B, Tramer MR. Complications of central venous catheters: internal jugular versus subclavian access--a systematic review. Crit Care Med 2002;30:454-60.
7. Vesely TM. Central venous catheter tip position: A continuing controversy. J Vasc Interv Radiol 2003;14:527-34.
8. Mayhall CG, Batt MD, Wong ES. Requirements for Infection Control Education and Practice Compliance Established for Physicians and Other Healthcare Workers in Minnesota. Infect Control Hosp Epidemiol 1993;14:503.‏
9. Mermel LA, Allon M, Bouza E, Craven DE, Flynn P, O'Grady NP, et al. Clinical practice guidelines for the diagnosis and management of intravascular catheter-related infection: 2009 Update by the Infectious Diseases Society of America. Clin Infect Dis 2009;49:1-45.
10. Kornbau C, Lee KC, Hughes GD, Firstenberg MS. Central line complications. Int J Crit Illn Inj Sci 2015;5:170-8.
Files
IssueVol 61 No 12 (2023) QRcode
SectionCase Report(s)
DOI https://doi.org/10.18502/acta.v61i12.16375
Keywords
Tricuspid valve Prosthetic malfunction Central venous pressure catheter

Rights and permissions
Creative Commons License This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.
How to Cite
1.
Behjati M, Tohidi F, Baharestani B. Malposition of Central Venous Pressure Catheter as a Rare Cause of Tricuspid Valve Prosthesis Malfunction: A Case Report malfunction; a case report. Acta Med Iran. 2024;61(12):770-772.